NovaSensor NPC-100 Series
Disposable Medical Pressure Sensors
The NovaSensor NPC-100 Series of Disposable Medical Pressure Sensors is specifically designed for use in disposable medical applications. The device is compensated and calibrated per the Association for the Advancement of Medical Instrumentation (AAMI) guidelines for industry acceptability. The sensor integrates a high-performance, pressure sensor die with temperature compensation circuitry and gel protection in a small, low-cost package.
The NPC-100 Series is manufactured in a class 1000 clean room to minimize possible sources of contamination. A specially designed, silicon, micromachined sensing element is used to meet or exceed all industry requirements, while minimizing assembly and test cost for maximum customer value. Thick-film laser trimming is employed for final compensation and calibration. Sensitivity is maintained to ±1% and linearity is better than 1% in the physiological operating pressure range.
The NPC-100 Series is batch-manufactured in ceramic plate form and shipped as an intact array or as single sensors packaged in Tape and Reel format for easy customer automation. This assembly method draws from well-established manufacturing techniques used in the electronics industry in order to produce a quality, high volume product.
- Medical instrumentation
- Blood pressure measurement
- Infusion pumps
- Kidney dialysis machines
- Solid state, high reliability
- Media compatibility
- High performance
- Factory filled with dielectric gel
- Small size
- Fully tested
- Temperature compensated
- Low cost disposable design
- Designed to AAMI specifications
General Parameters | Value | Units | Notes |
Pressure Range | -30 to 300 | mmHg | -0.58 psi to 5.8 psi (-0.03 bar to 0.03 bar) |
Overpressure | 125 | psi | (x bar) minimum |
Electrical Parameters* | Value | Units | Notes |
Input Excitation | 1 to 10 | VDC | Calibrated for 6 VDC |
Dielectric Breakdown | 10,000 | VDC | 5 |
Risk Current | 2 | µA | Maximum (per AAMI), 5 |
Input Impedance | 285 to 315 | Ω | |
Output Impedance | 285 to 315 |
Ω |
|
* Electrical @ 72°F (22°C), unless otherwise stated. |
Environmental Parameters | Value | Units | Notes |
Compensated Temperature | 59°F to 104°F | °F | (15°C to 40°C) |
Operating Temperature | 59°F to 104°F | °F | (15°C to 40°C) |
Storage Temperature | -13°F to 158°F | °F | (-25°C to 70°C) |
Humidity | 10 to 90 | % | |
Light Sensitivity | 1 | mmHg | maximum (per AAMI BP22) |
Operating Product Life | 168 | hours | |
Shelf Life | 3 | years |
Mechanical Parameters | Value | Units | Notes |
Weight | <0.0044 | lb | (<2 g) |
Volume Displacement | <0.0008 | in3 | (<0.02 mm3) |
Media Interface | Medical grade, dielectric gel | ||
Gel Tube Interface Material | Polycarbonate |
Performance Parameters1 | Value | Min | Typ | Max | Notes |
Offset | mmHg | -25 | 0 | 25 | |
Sensitivity | V/V/mmHg | 4.95 | 5 | 5.05 | |
Calibration | mmHg | 97.5 | 100 | 102.5 | 2 |
Symmetry | % | ±5 | |||
Linearity (-30 to 100 mmHg) | mmHg | 1 | 6 | ||
Linearity (100 to 200 mmHg) | % output | 1 | 6 | ||
Linearity (200 to 300 mmHg) | % output | 1.5 | 6 | ||
Thermal Coefficient Offset | mmHg/°C | ±0.3 | 3 | ||
Thermal Coefficient Span | %/°C | ±0.1 | 3 | ||
Frequency Response | Hz | 1200 | 5 | ||
Phase Shift | degrees | 5 | 5 | ||
Offset Drift | mmHg/8 hrs | 1 | 4, 5 | ||
|
- NCP-100
- NPC-100-T